Video summary
Pressure Transmitter Workshop Re-range - [HART 475 / Fluke 789 / Druck DPI610 / Foxborough TX]
Main summary
Key takeaways
Pressure transmitter re-range and calibration check (tutorial)
Purpose
Demonstrates re-ranging a Foxborough IG10P pressure transmitter from 0–10 bar to 0–5 bar and performing a 5‑point calibration check.
Test equipment used
- Druck DPI 610 IS — pressure source/calibrator (pressure application and pump)
- Fluke 789 ProcessMeter — provides 24 V loop power and reads mA output
- HART 475 field communicator — HART configuration and send-to-instrument
- Mechanical adapters/tubing — 1/2” NPT to 1/4” Gyrolok adapter, PTFE tape, small‑bore tubing and olive fitting
- Paper calibration/check sheet (5‑point)
Procedure (high level)
-
Mechanical connections
- Removed transmitter back plate and wired two leads to the Fluke 789.
- Fitted a 1/2” NPT → 1/4” Gyrolok adapter; used PTFE tape on the NPT threads and tightened with wrenches.
- Connected Gyrolok tubing to the Druck DPI 610.
-
Pressure application and leak check
- Pressurized with the DPI (air used in demo; hydraulic pump recommended for higher pressures).
- Held at 5 bar briefly to confirm no significant leak, then vented.
-
Loop powering and initial readings
- Fluke 789 set to loop‑power (mA source); red to positive, black to negative.
- Observed ~3.99–4.00 mA at atmospheric pressure (expected 4.00 mA for 0 bar on a 0–10 bar range).
-
HART re-range (PV rearrange)
- Connected HART 475 to HART pins and went online.
- Confirmed LR = 0 bar and URV = 10 bar.
- Navigated: Device setup → Calibrate → PV rearrange → edited URV to 5 bar → pressed Send to write to instrument.
- Returned the loop to auto and disconnected the HART communicator.
-
5‑point calibration check
- Applied pressures: 0.00, 1.25 (25%), 2.50 (50%), 3.75 (75%), 5.00 (100%) bar.
- Technique:
- Use coarse then fine adjust on the DPI; approach setpoints from below to minimize hysteresis.
- Allow settling time (transmitter filtering can slow response).
- Recorded mA readings (rounded to two decimals). Typical measured errors were around −0.01 to −0.02 mA.
-
Shutdown
- Powered down equipment and removed connections; transmitter is ready for field use.
Error calculation and results
- Span = 20 mA − 4 mA = 16 mA
- Percent error = (mA error / 16 mA) × 100
-
Example: a 0.01 mA error → (0.01 / 16) × 100 = 0.06% a 0.02 mA error → (0.02 / 16) × 100 = 0.13%
-
These are very small errors and are within acceptable calibration limits shown in the tutorial.
Note: The HART communicator shows a star next to URV/LRV when changes are staged in the communicator but not yet written — you must press Send to commit the changes to the instrument.
Practical tips and notes
- Use PTFE tape on NPT threads to prevent leaks.
- Use a hydraulic pump for higher pressures (air is slow/inefficient for high pressures).
- Avoid overshooting setpoints and then coming back down; approach setpoints from below to reduce hysteresis effects.
- Use the fine adjust on the pressure source for precise setpoints.
- Allow time to settle because transmitter filtering can slow the response.
- The DPI could supply loop power, but the demo used the Fluke 789 for loop power and mA readout.
Devices/brands referenced
- Foxborough IG10P (pressure transmitter)
- Druck DPI 610 IS (pressure calibrator)
- Fluke 789 ProcessMeter (loop power and mA measurement)
- HART 475 field communicator (HART device configuration)
Main speaker / source
- Tutorial instructor / demonstrator (unnamed narrator in the video)
- Equipment manufacturers: Foxborough (transmitter), Druck (DPI 610), Fluke (789), and HART (475 communicator)
Rate this summary
Your feedback will help improve summaries.
Improve this summary
Reprocess with a stronger model when the summary feels incomplete or inaccurate.
Translate summary in another language
Ask questions to this video
Chat for follow-up questions, clarifications, and source-backed answers.