Summary of "How To Perform DNA Quantitation Using a Spectrophotometer"
Summary of “How To Perform DNA Quantitation Using a Spectrophotometer“
This video provides a step-by-step guide on how to quantify DNA concentration using a spectrophotometer, specifically focusing on measuring purified plasmid DNA.
Main Ideas and Concepts
- Purpose: Estimate the amount and purity of DNA in a purified plasmid preparation using a spectrophotometer.
- Instrument Preparation: Turn on the spectrophotometer about 10 minutes before starting to allow it to warm up.
- Sample Handling: Use UV-transparent cuvettes and handle them carefully by holding only the opaque sides to avoid contamination or damage.
- Measurement Principle: DNA concentration is determined by measuring absorbance at 260 nm, and purity is assessed by the ratio of absorbance at 260 nm to 280 nm.
Detailed Methodology / Instructions
-
Preparation:
- Turn on the spectrophotometer approximately 10 minutes before use.
- Obtain two UV-transparent cuvettes.
- Label one cuvette as “C” (control) and the other as “K” (test/sample).
-
Sample Preparation:
- Set micropipette to 245 µL.
- Pipette 245 µL of distilled water into both cuvettes.
- Set micropipette to 5 µL.
- Add 5 µL of purified DNA sample into the test cuvette (K).
- Mix the sample by pipetting up and down gently.
-
Spectrophotometer Setup:
- Set the spectrophotometer to the DNA quantitation mode.
- Confirm the screen shows DNA selected.
- Press enter to confirm readiness for absorbance reading.
-
Blank Measurement:
- Insert the control cuvette (C) with distilled water, ensuring correct orientation.
- Press the “read blank” button.
- The screen should show zero absorbance.
-
Sample Measurement:
- Remove the control cuvette.
- Insert the test cuvette containing the DNA sample.
- Press the “read sample” button.
- Press the right-arrow button to view results.
-
Reading Results:
- Absorbance value will be displayed (e.g., 0.011).
- Concentration will be displayed in µg/mL (e.g., 0.5415 µg/mL).
- Multiply the concentration by the dilution factor (e.g., 50) to get the final DNA concentration.
-
Purity Assessment:
- Press the button to display the 260/280 absorbance ratio.
- A ratio greater than 1.8 indicates pure DNA.
- Example result: 5.49, indicating highly pure plasmid DNA.
Key Lessons
- Proper cuvette handling is crucial to avoid contamination.
- The blank measurement with distilled water is essential for calibration.
- DNA concentration must be adjusted by the dilution factor.
- The 260/280 ratio is a standard metric to assess DNA purity.
- Spectrophotometry is a quick and reliable method for DNA quantitation.
Speakers / Sources
The video appears to feature a single narrator/instructor guiding through the procedure. No other speakers or external sources are explicitly identified.
Category
Educational
Share this summary
Featured Products
Spectrophotometer, Digital Visible Laboratory Automatic Spectrometer, Visible Wavelength Range 350-1020nm Wave Mass Spectrometer with Automatic Light Door for R&D Teaching Experimentation
BrandTech Semi-Micro UV-Transparent Disposable Cuvettes, 1.5mL to 3.0mL Sample Volume, 4.5 X 23mm Window Dimensions, 230-900nm wavelength. 100/Pack
Micropipette 1000-5000ul Adjustable Manual Volume Micropipette M5000 for Laboratory, Medical Sampling Testing
Forensic Biology: Identification and DNA Analysis of Biological Evidence
50Pcs 2mL Spin Column Micro centrifuge Tube, High Speed Plasmid Column & Collection Tube Kit by Biomed Scientific