Western Blotting

Main summary

Key takeaways

Educational

Summary

The video demonstrates how to perform a Western blot to detect specific proteins in a sample. The technique can also provide information about a protein’s approximate size and relative abundance.

Procedure

1. Prepare the gel and materials

  • Fill a tray with blotting buffer.
  • Open the gel cassette using the opening key, aligning the arrows on the lever and cassette.
  • Trim the top and bottom edges of the gel, then equilibrate it in blotting buffer for 15 minutes on a rocking platform.
  • Thoroughly soak two fiber pads in blotting buffer.

2. Assemble the blotting sandwich

  1. Partially fill a suitable container with blotting buffer to a depth of about 1 cm.
  2. Place the gel-holder cassette in the container with its black side down and immersed in buffer. Keep the white side above the buffer.
  3. Layer the following items on the black side:
    • A fiber pad
    • A piece of blotting paper
    • The equilibrated gel
    • A wet nitrocellulose membrane
    • A second piece of wet blotting paper
    • A second fiber pad

Avoid trapping air bubbles between layers. Once the membrane contacts the gel, avoid shifting it, because protein transfer begins immediately. Use a roller to remove bubbles between the gel and membrane.

Fold the clear side of the holder over the sandwich and secure it with the white clip.

3. Transfer proteins to the membrane

  • Insert the holder into the inner module, matching the black sides, then place the module in the electrophoresis chamber.
  • Add a frozen cooling unit and fill the chamber with blotting buffer to the level of the holder’s white clip.
  • Close the chamber and connect the leads correctly: red to red, black to black.
  • Run the blot at 20 volts for 2.5 hours.
  • Turn off the power before disassembling the apparatus.

4. Check the transfer

Open the holder in a container of blotting buffer, black side down, and remove the layers until you reach the nitrocellulose membrane. The proteins should have transferred from the gel to the membrane.

The video notes that prestained Kaleidoscope standards are visible on both sides of the membrane and that no proteins remain visible on the gel.

5. Detect the target protein

  1. Block the membrane in 25 mL of blocking solution for 15 minutes at room temperature on a rocking platform.
  2. Pour off the blocking solution and add 10 mL of primary antibody. Incubate for 10–20 minutes with fast rocking to keep the membrane continuously covered.
  3. Pour off the primary antibody. Rinse quickly with 50 mL of wash buffer and discard it, then wash in a fresh 50 mL of wash buffer for 3 minutes at medium speed.
  4. Discard the wash buffer, add 10 mL of secondary antibody, and incubate for 5–15 minutes at fast speed.
  5. Pour off the secondary antibody, rinse quickly with 50 mL of wash buffer, then wash with a fresh 50 mL for 3 minutes at medium speed.
  6. Add 10 mL of substrate and incubate for 10–30 minutes, either with manual shaking or on the rocking platform. Monitor color development.
  7. Once color develops, rinse the membrane twice with distilled water, blot it dry, air-dry it for 3–60 minutes, and cover it with plastic wrap for storage.

Speakers and sources

  • Narrator: An instructional voice explains the Western blot procedure; no individual speaker is identified.
  • Source: Bio-Rad Laboratories instructional video.

Rate this summary

Your feedback will help improve summaries.